Unannounced X-Ray Inspections

To improve operational efficiency, the Bureau of Radiological Health is transitioning to new inspection procedures which include performing unannounced inspections.

Regulation 61-64 X-Rays (Title B) Section 10.5 requires each registrant afford Department inspectors opportunity to inspect machines, activities, facilities, premises, and records at all reasonable times. Section 1.10 requires records to be readily available at the facility for Department review.

Shielding Plan Requirements

Shielding Plans Are Required For:

  1. All diagnostic medical X-ray units installed at new facilities.
  2. Relocation of a unit within a facility.
  3. All installations of therapeutic X-ray or accelerator units.
  4. All installations of medical X-ray units at veterinary offices.
  5. All industrial X-ray units used in shielded room configurations.
  6. All dental cephalometric, dental CTs and TMJ units.
  7. Replacement of an existing X-ray unit at a facility, to include the control, and/or generator.
  8. Table bone density and mammography units (or are